Description
Surtron LED
SURTRON 400 HP can be used for simultaneous monopolar and bipolar surgery in applications requiring high precision and reliability.
UNDERWATER CUTTING AND COAGULATION
DUAL ACTIVITY
– Two output channels for monopolar cutting
– Two independent setting and output channels for more frequent monopolar coagulation.
CONTROLLED CUTTING PROCEDURES
– Cutting currents for an eschar-free incision
– Cutting current with selectable coagulant property for bleeding-controlled dissection
– Low capacitive coupling by current endoscopic procedures Slow BLEND
DOUBLE SUBSIDIARY COAGULATION PROCEDURES
– SPEEDY Coag DEEP ** * and two operators
– SPRAY COAG can be performed using a handpiece or foot pedal
WIDE BIPOLAR CHARACTERISTICS
– Bipolar cut with bleeding control
– Slow BLEND cutting current
– Automatic start and stop activated separately
TOTAL SAFETY FOR PATIENT AND SURGEON
– The acceptability of the contact impedance between the patient’s body and the neutral plate is constantly checked by the SPEC electronic skin plate control.
– Continuous monitoring of leakage current
– Automatically compensated mains voltage change before RF power start-up
CUSTOMISED PROCEDURES
– Independent choice of coagulation currents and power levels for each of the two operators
– Ten personal procedures that can be configured and immediately recalled
CONTROL OF EXTENSIONS AND CONNECTION OF SIGNALS
– outputs for connection of controlled argon plasma units and smoke extractor
AUTOMATIC SELF-DIAGNOSIS WITH SURTRON 400 HP
– Automatic check of the functionality of the complete electronic drive and connected accessories and coded report of detected faults.
* Coag SPEEDY is also known as Fulgurate or Forcé.
** Coag DEEP is also known as Pin Point, Dessicate or Soft.
SPECIFICATIONS
Code 10400.902
Maximum monopolar pure CUT output power: 400 W – 300 Ω
Maximum monopolar cut-coag CUT80% output power: 300 W – 300 Ω
Max. monopole cut-coag output power CUT60%: 250 W – 300 Ω
Maximum monopole output power CUT40%: 200 W – 300 Ω
Maximum monopole cut-coag output power ENHANCED: 250 W – 500 Ω
Maximum monopole output power coag cut BLEND: 250 W – 300 Ω
SPEEDY coag monopolar maximum output power: 120 W – 500 Ω
Maximum power output of monopolar coagulator DEEP: 120 W – 200 Ω
Maximum output power of SPRAY monopolar coagulator: 120 W – 2000 Ω
Maximum output power of SPRAY ARGON monopolar coagulator: 120 W – 2000 Ω
Maximum bipolar CUT output power: 120 W – 50 Ω
Maximum bipolar cut-coag CUT80% output power: 120 W – 50 Ω
Maximum bipolar cut-coag CUT60% output power: 120 W – 50 Ω
Maximum bipolar cut-coag output power CUT40%: 100 W – 50 Ω
Maximum bipolar cut-coag output power BLEND: 120 W – 150 Ω
Maximum output power bipolar coag COAG: 120 W – 50 Ω
Tank sealing: 200 W – 50 Ω
Working frequency Monopolar / Bipolar: 425 kHz / 525 kHz
Patient circuit: F
Selectable input voltage: 115 – 230 Vac
Mains frequency: 50 – 60 Hz
Electrical input power: 1000 VA
Dimensions WxHxD: 470x150x400 mm
Weight: 17.5 kg CONTROLS
Patient/plate circuit control
Output power control
Self-check control
H.F. leakage current control
Mains fluctuation compensationSECURITY
EN60601-1
EN60601-1-2
EN60601-2-2
Enclosure protection level: IP21
Electrical classification: CF
MDR 2017/745/EU Class: II b
Complies with safety standards: EN 60601-1, EN 60601-1-2; EN 60601-2-2. Electrical classification: I CF.
Réf. IBSUR100